Sentrac Strix Hydrogen Leak Detectors
- High sensitivity
- Low maintenance costs
- Wide measurement ranges
- Fast recovery time
- By using diluted hydrogen gas (95 % nitrogen, 5 % hydrogen), minute leaks can be detected.
Using inexpensive general industrial gas (95 % nitrogen, 5 % hydrogen) as a tracer gas and a sensor that does not react to gases other than hydrogen gas realizes highly sensitive measurement.
Even immediately after detecting a large leak, sensitive measurement can be made immediately.
Basic Specs.
Sentac (Desktop model)
Minimum detectable leak: P60 hand probe Locating mode Measurement mode |
1×10-8 Pa・m3 /sec (with 5 % hydrogen) 0.5 ppm or 5×10-8 Pa・m3 /sec (with 5 % hydrogen) |
---|---|
Rising time | Approx. 1 minutes |
Calibration method | Leak calibrator or 10 ppm calibration gas with a balance of air |
Interface |
D-sub 25 pins RS-232C Audio output Analog signal output Digital signal (3-input/4-output) Mini USB SD card reader |
Power source | 100 to 240 VAC, 50/60Hz, 2A |
Main unit dimensions (W x H x D) | 305 x 165 x 182 mm |
Main unit weight (excluding probe and cable) |
4.2kg |
Sentac (portable model)
Minimum detectable leak: P60 hand probe Locating mode Measurement mode |
1×10-8 Pa・m3 /sec (with 5 % hydrogen) 0.5 ppm or 5×10-8 Pa・m3 /sec (with 5 % hydrogen) |
---|---|
Rising time | Approx. 1 minutes |
Calibration method | Leak calibrator or 10 ppm calibration gas with a balance of air |
Interface |
D-sub 25 pins RS-232C Audio output Analog signal output Digital signal (3-input/4-output) Mini USB SD card reader |
Power source | 100 to 240 VAC, 50/60 Hz, 0.3 A Rechargeable lithium-ion battery |
Main unit dimensions (W x H x D) | 330 x 200 x 280 mm (including case) |
Main unit weight (excluding probe and cable) |
4.2 kg |
Operation time | 12 hours (without screen saver) |
Charging time | 6.5 hours |
Features / Functions
Features of diluted hydrogen gas
- Hydrogen gas is not easily affected by the atmosphere because it is present at only 0.5 ppm in the air.
- It is cheaper than helium gas.
- A mixture of 5 % hydrogen and 95 % nitrogen is not flammable.
- Since it is not a natural resource, a stable supply is possible.
Features of the product
- It does not react to gases other than hydrogen, so it is not affected by other gases in the measurement environment.
- The sensor at the tip of the probe reacts to hydrogen in a short time.
- Use the OLED on the hand probe to check the leak and other information without having to check the main unit.
- Wide measurement range for large and small leaks.
- It will recover quickly even after detecting a large leak.
- The portable model does not require a power supply and can be used in the field.
- Can be used for leak testing not only in manual operation but also in automated systems.
Main applications
- Automotive industry
- Aerospace industry
- Refrigeration and air conditioning industry