FM-2C Peripheral
- Flow Master FM-2C is ideal for flow measurement and daily inspection of flow testers
- FM-2C is a standard flowmeter with a fixed pressure and flow.
Unlike float-type flowmeters, there is no need to determine the flow using a regulator valve and no trouble such as damage to the float due to inrush pressure.
Basic Specs.
FM-2C Production range
Customized to the desired flow. Use it as a reference instrument.
When placing an order, specify the test pressure and flow in the following ranges.
Test pressure | Flow | |
---|---|---|
Vacuum | -89 to -50.1 kPa | 200 mL/min to 5 L/min |
-50 to -21.1 kPa | 100 mL/min to 15 L/min | |
-21 to -10.1 kPa | 100 mL/min to 10 L/min | |
-10 to -1 kPa | 50 mL/min to 3 L/min | |
Pressure | 1 to 10 kPa | 50 mL/min to 3 L/min |
10.1 to 21 kPa | 100 mL/min to 10 L/min | |
21.1 to 51 kPa | 200 mL/min to 15 L/min | |
51.1 to 100 kPa | 0.3 L/min to 25 L/min | |
100.1 to 150 kPa | 0.5 L/min to 35 L/min | |
150.1 to 200 kPa | 0.5 L/min to 45 L/min | |
201 to 250 kPa | 0.5 L/min to 55 L/min | |
251 to 300 kPa | 0.5 L/min to 65 L/min | |
301 to 350 kPa | 0.5 L/min to 70 L/min | |
351 to 400 kPa | 0.5 L/min to 80 L/min |
For specifications outside the production ranges, a request for quotation (RFQ) will be required.
201 to 209 mL/min is outside the standard production range.
2.01 to 2.09 L/min is outside the standard production range.
Test pressure | Range | Minimum unit |
---|---|---|
Vacuum | Down to -89 kPa | 0.1 |
Pressure | Less than 200 kPa | 0.1 |
200 kPa or more | 1 |
Required flow | Difference from the required flow | Minimum unit |
---|---|---|
50 mL to 200 mL | Accuracy (Difference from the required flow): Within ±10% | 1 mL |
0.21 L to 2.00 L | 0.01 L | |
2.1 L to 80 L | 0.1 L |
For the flow above 200 mL/min, specify in units of L/min.
Main body: FM-2C
Accessories: (IN-side and OUT-side joints (1 set), O-ring (3 pcs.)
Model FM-2C(AB)(C,D)
Example:
HIGH-side joint:R1/2, LOW-side joint:Ø8 tube, Required flow:200 mL/min, Test pressure:10 kPa
FM-2C(R32)(200 mL/min, 10 kPa)
A, B, C and D in the model notation are mandatory.
A | HIGH-side joint | R2 | Port size R1/4 |
---|---|---|---|
R3 | Port size R1/2 | ||
R4 | Port size Ø8 tube | ||
B | LOW-side joint | Port size R1/4*1 | |
1 | Port size R1/2 | ||
2 | Port size Ø8 tube | ||
C | Required flow | Please check the production range. | |
D | Test pressure |
*1 the joint connection diameter of the LOW side is R1/4, leave the symbol of LOW joint blank.
Features / Functions
Features
Customized to the specified pressure and flow.Use as a standard for testing and comparing flowmeters.
- Improved long-term flow stability.
- Enhanced prevention of distortion during installation.
- Check the sensitivity of flow testers easily and reliably.
- Extremely low flow variation ensures long-time stable use.
- Use as a flow master for measuring deviation flow. (Test for ventilation and bore diameter of holes)
- 15% more compact than conventional products.
- The numerical value of the tester can be checked by connecting to the CAL port of the tester.
- Can be used in Flow Check (C-CHK), (F-CHK) of AF series.
- Joints for connecting to the piping available.
- Easy connection and disconnection with a plug.
- ISO/IEC17025 calibration available.
Example of use with our Air Flow Tester