Basic Specs.
Flow range | Two ranges can be selected from 5, 20, 100, and 200 mL/min. |
---|---|
Flows |
5 mL/min: 0.00 to 5.00 20 mL/min: 0.00 to 19.99 100 mL/min: 0.00 to 100.0 200 mL/min: 0.00 to 199.9 |
Test pressure range |
L: 5 to 100 kPa M: 0.005 to 0.7 MPa |
Test pressures |
L: 05.0 to 100 (kPa) M: 0.005 to 0.700 (MPa) |
Test fluid | Air |
Linearity accuracy | ±2% of F.S. ±1digit (at calibration temperature) |
Guaranteed temperature range and accuracy | ±3.5% of F.S. (at 10 to 30℃) |
Operating temperature | 5 to 40℃ |
Air source | Dry air (max. 990 kPa) |
Power source | 100 to 240 VAC ±10% |
Power cable |
VA: 125 VAC, 3 m VE: 250 VAC, 2 m VK: 250 VAC, 2 m (for Chinese customers) |
Features / Functions
The flow range can be changed.
- Up to 2 ranges can be selected from 4 ranges.
- The flow sensor range can be changed on the front panel.
- By changing the ranges, a wide range of flows can be covered to manage many Leak Masters.
Easy to set test pressure
- The precision regulator allows you to set the desired pressure precisely.
- Test pressure can be reliably checked with the digital pressure gauge.
- Pressure ranges for low pressure (L) and medium pressure (M) are available.
Documents