QLC Series / LC Series Peripheral
- Manual Leak Calibrator
- Manual Leak Calibrator is for differential pressure Air Leak Testers that can generate desired volumetric changes.
Easily check the sensitivity of Air Leak Testers and measure the Equivalent Internal Volume.
Basic Specs.
Calibrator Specifications
Name | Model | Maximum volumetric change |
Operating pressure range (Recommended) |
Notes |
---|---|---|---|---|
ALC Series | ALC-05 | 0.5 mL | 0 to 4.9 MPa | Built into Air Leak Testers. K(Ve) Automatic Setup is available. Can be used for K(Ve) Check. Pilot pressure: 0.4 to 0.7 MPa |
ALC-1 | 1 mL | 0 to 2 MPa | ||
ALC-4 | 4 mL | 0 to 0.8 MPa | ||
ALC-10 | 10 mL | 0 to 0.2 MPa | ||
QLC Series | QLC-0021 | 0.2 mL | 0 to 0.5 MPa | It generates predetermined volumetric change spontaneously by simply pushing a button. Can be used even when Detection time is short. |
QLC-0101 | 1 mL | 0 to 0.5 MPa | ||
QLC-0401 | 4 mL | 0 to 0.5 MPa | ||
QLC-1001 | 10 mL | 0 to 0.2 MPa | ||
LC Series | LC-11 | 1 mL | 0 to 0.4 MPa | Turn the dial in Detection to cause volumetric change. If Detection time is short, the time has be extended. |
LC-12 | 1 mL | 0 to 0.4 MPa | ||
LC-22 | 2 mL | 0 to 0.2 MPa | ||
LC-42 | 4 mL | 0 to 0.1 MPa |
Features / Functions
QLC (Quick Leak Calibrator)
It generates predetermined volumetric change spontaneously simply by pushing a button.LC (Leak Calibrator)
It is a cylinder type manual leak calibrator that can give an accurate volumetric change specified on the scale.